Neoventa
This is the US version of the site. Change region...

Tech Specs

Overview Features Accessories Tech Specs  

STAN® Main Unit

(Display Unit and PIB mounted together)

Physical Characteristics

  • Dimensions (width x depth x height): 16.5 x 5.6 x 13.1 in.
  • Weight: 20 lbs

Power

  • Operating voltage: 100-240 V
  • Line frequency: 50/60 Hz
  • Power consumption: 90 VA max, 35 W typ.

Operation environment

  • Operating temperature: +50°F to +104°F
  • Relative humidity: 30% to 75% non-condensing
  • Atmospheric pressure range: 700hPa to 1060 hPa

Transport and storage environment

  • Transport (inside packing) or storage (without packing)
  • Temperature range: -13°F to +158°F
  • Temperature range (max 16 hours): -40°F to +158°F
  • Relative humidity range: 10 % to 90 %, no condensation
  • Relative humidity range (max 16 h): 10 % to 95 %, incl. condensation
  • Atmospheric pressure range: 500 hPa to 1060 hPa

Display

15" 1024x768 TFT LCD Touch screen, with tilt and rotation and brightness control, displaying the following output data:

  • Dual Heart Rate: 50-240 bpm (US), 30-240 bpm (FECG)
  • T/QRS ratio: -0.30 to +0.90 (FECG)
  • Fetal ECG Average with biphasic (BP) ST indication
  • Uterine Activity: 0-100 units (external), 0-100 mmHg (internal)
  • Dual heart beat and signal quality indications
  • Scrolling traces (HR: 50-210 or 30-240 bpm, T/QRS: -0.12 to +0.50, UA: 0-100, BP, time):
  • 1, 2 or 3 cm/min, selectable range 5 to 30 minutes live data, scrollable window (up to 25 minutes) over complete recording
  • Event Log information: FECG waveform data, operator notes, signal quality
  • Realtime FECG signal
  • Signal and system status
  • Patient name and id
  • Easy-to-use touch screen control fields, with expandable control functionality

Audible Indications

  • FECG-generated heart beat beep
  • US-generated heart sounds

Digital Data Storage

  • Integrated flashdisk automatically storing 3000 hours of latest recorded data
  • USB Connector for removable USB storage devices, or transfer of digital data to archiving systems (optional)

Recording

The following applies when standard transducers are used.

Fetal ECG Recording

  • Maximum electrode potential difference: ±0.75 VDC
  • Input range: ±10 mV
  • Pre-amplifier bandwidth: 0.05 to 100 Hz (-3dB)
  • Sensitivity FHR detection, minimum QRS ampl. 50 μV p-p
  • Sensitivity ECG analysis, minimum QRS ampl. 200 μV p-p
  • Mains frequency rejection: >40dB
  • Beat-to-beat HR detection
  • HR range: 30-240 bpm
  • HR accuracy: ±1 bpm (up to 180 bpm), ±2 bpm (above 180 bpm)
  • 30-beat ECG Averages.
  • ECG average T/QRS ratio range: -0.30 to +0.90
  • T/QRS accuracy: ±0.02
  • Bi phasic ST segment indication: Grade 0, 1, 2 and 3

US Recording

  • Centre frequency: 1.0 MHz (US2) and 1.5 MHz (US1)
  • Intensity: <20mW/cm2
  • HR range: 50-240 bpm
  • HR accuracy: ±2 bpm (up to 180 bpm) ;±3 bpm (above 180 bpm)

TOCO Recording

  • Max load: 500 g
  • Output range: 0-100 units
  • Sensitivity: 0.4 units/g (nominal, adjustable)
  • Manual output offset: 5 units
  • Resolution: 1 unit
  • Accuracy: ±20% of display
  • Bandwidth: DC to 0.15 Hz (-2dB) (-40 dB at 1Hz)

IUP Recording

  • Range: 0-100 mmHg (0-13.33 kPa)
  • Resolution: 1 mmHg
  • Accuracy: ±5% of display
  • Bandwidth: DC to 0.15 Hz (-2dB) (-40 dB at 0.5Hz)

Trolley STAN TR31

Physical Characteristics

  • Dimensions (width x depth x height): 22 x 23 x 55 in
  • Weight: 46 lbs
  • Input mains cable: Removable

Power

  • Nominal operating voltage: 100-240 V
  • Line frequency: 50/60 Hz
  • Max load 10 A

Classification

STAN S31 fetal heart monitor is an FDA approved class III medical device.